Model
|
IJ201
|
IJ300
|
Method
|
Forced convection
|
Temperature control
range (*1)
|
0 to 60°C (Ambient temp.: 20°C, No load)
|
Temp. adjustment accuracy (*1)
|
±0.5°C ( at 37°C)
|
Temp. distribution accuracy (*1)
|
±1.0°C ( at 37°C)
|
Temp. rise time (*1)
|
Approx. 50min. (20°C to 60°C)
|
Approx. 60min. (20°C to 60°C)
|
Temp. fall time (*1)
|
Approx. 120min. (20 to 0°C)
|
Heater
|
Pipe heater 150w
|
Pipe heater 300w
|
Cooling device
|
2peltiert devices, forced radiating method
|
4peltiert devices, forced radiating method
|
Itworks continuous at the setting temperature below 35°C.
(It is turned off at the setting temperature 35°Cand over.)
|
Internal fan
|
Axial fan (DC)
|
Controller
|
VS4 type
|
Temperature control
system
|
PID control for heater output by microprocessor
|
Setting system
|
Digital setting by menu key and up/down keys
|
Operation mode
|
Fixed temperature, Quick auto stop, Auto stop,Autostart,Program operation
|
Sensor
|
Temperature sensor: Platinum resistance (Pt 100Ω)
Overheating prevention sensor: K-thermocouple
|
Additional functions
|
Lock function, Auto recovering afterpowerfailure,Calibration offset
|
Self-diagnostic
functions
|
Failure of Sensor, heater, SSR,automaticoverheatingprevention device
|
Safety device
|
Earth leakage breaker, Overheating preventiondevice
|
External dimensions
(W×D×H mm) (*2)
|
580×417×426
|
450×496×665
|
Internal dimensions
(W×D×H mm)
|
300×300×300
(Air jacket installed: 250×250×250 mm)
|
350×350×350
(Air jacket installed: 300×300×300 mm)
|
Capacity
|
27L (Air jacket installed: 15.6L)
|
43L (Air jacket installed: 27L)
|
Door
|
Single door with magnetic packing
|
Single door with magnetic packing
|
Power source
(50/60Hz)
|
100V AC 4.5A / 220V AC 2.1A
|
100V AC 8A / 220V AC 4A
|
Weight
|
Approx. 25Kg
|
Approx. 37Kg
|
Accessories
|
2 Shelves (withstand load: 10kg/one shelf),Platefordrainage, Instruction manual
|
Optional accessories
|
Extra shelf (with bracket), Air jacket madeofaluminum(with shelf),
Inner door made of acrylic resin
|